To calibrate, click Analysis >> Analysis/Single Level Calibration.

8

The Analysis/Single Level Calibration window will display.

9

In the Sample ID: box, type the sample identification, if not already present.

10

In the Method: box, select the name of the method to be calibrated, if not already present. To
do so, click the Open file icon adjacent to the field.

11

In the Data path: box, select the data path name, if not already present. To do so, click the
Open file icon adjacent to the field.

12

In the Data file: box, select the file name, if not already present. To do so, click the Open file
icon adjacent to the field.

13

The Amount values will remain set at 1; a value of 1 corresponds to a single level calibration.

14

Select Calibrate. In the Calibration Level box, type 1.

NOTE: For the initial calibration, it is not necessary to select any of the boxes involving

clearing calibrations or replicates. However, if the method was previously calibrated
and the calibrated information needs to be replaced, select Clear all calibration
before starting.

Table 13-1 Conduct a single level calibration (continued)