Changing the Instrument IP Address

Table 3-1 Changing the Instrument IP Address

Step

Description

1

The IP address of the instrument can be changed through the embedded web page.

NOTE: Consult an IT professional to assist with IP settings.

2

Open a web browser, such as Internet Explorer.

3

In the address bar, type the current IP Address of the instrument. Press Enter.

NOTE: The factory default IP address is set at 10.1.1.101.

4

The instrument’s embedded web page will display. Click the IP Configuration tab. The current
IP address for the instrument is displayed. Click Make Changes.

5

When prompted for a User name and Password, type ipconfig for both. Click OK.