Click (Windows) Start >> Run.

14

In the Run window, in the Open: box, type ping 192.168.1.99. Click OK.

15

Verify from the Command Prompt that the PC is receiving data packets from the instrument.

NOTE: If the command prompt reads Request Timed Out or Detection Limit Unreachable,

repeat steps 3 to 7 to verify IP settings. Verify that 3000 Micro GC is on connected to
the PC via the crossover cable. If there is still no communication contact INFICON.

16

Open a web browser, such as Internet Explorer.

17

In the address bar, type 192.168.1.99 and press Enter.

18

The instrument embedded web page will display. Select the IP Configuration tab.
Review the IP address.

Table 2-1 Connecting to the instrument via the static IP address (continued)

Step

Description