15

When the analysis is complete, the chromatogram will be integrated. The areas for the peaks
identified as calibration compounds will be entered into the method. In addition, a calibration
curve will be generated using these areas. The method is now calibrated for one of the
desired levels.

16

Repeat steps 4 to 15 to calibrate other levels, one level at a time.

17

After completing a calibration in the window, click Method >> Review Calibration to review
the calibration curve and its associated data. Each peak will have its own calibration curve.
If desired, right-click on a blank area to bring up options such as View Fit Type, Change Fit
Type
, and a Concentration Calculator.

Table 14-1 Conduct a multiple level calibration (continued)

Step

Description