Guide 4 changing the carrier gas configuration, 3000 micro gc guide 4 – INFICON 3000 Micro GC Gas Analyzer Reference Guide User Manual

Page 23

Advertising
background image

4 - 1

PN

07

4-

55

8-

P1

A

3000 Micro GC Guides

3000 Micro GC Guide 4

Changing the Carrier Gas Configuration

Table 4-1 Changing the carrier gas configuration

Step

Description

1

The type of carrier gas used for analysis can be changed through the embedded web page.

NOTE: If a carrier gas type is changed without following this procedure, inverted peaks will

appear in the chromatogram.

2

Open a web browser, such as Internet Explorer.

3

In the address bar, type in the current 3000 Micro GC IP Address. Press Enter.

NOTE: At the factory, the instrument IP address was set to 10.1.1.101.

4

The instrument’s embedded webpage will display. Click the Gas Type tab. The current gas
configuration for installed GC modules is shown. Click Make Changes.

5

When prompted for a User name and Password, type gasconfig for both.