Changing the Carrier Gas Configuration
Table 4-1 Changing the carrier gas configuration
Step
Description
1
The type of carrier gas used for analysis can be changed through the embedded web page.
NOTE: If a carrier gas type is changed without following this procedure, inverted peaks will
appear in the chromatogram.
2
Open a web browser, such as Internet Explorer.
3
In the address bar, type in the current 3000 Micro GC IP Address. Press Enter.
NOTE: At the factory, the instrument IP address was set to 10.1.1.101.
4
The instrument’s embedded webpage will display. Click the Gas Type tab. The current gas
configuration for installed GC modules is shown. Click Make Changes.
5
When prompted for a User name and Password, type gasconfig for both.