INFICON 3000 Micro GC Gas Analyzer Reference Guide User Manual
Page 34
7 - 2
PN
07
4-
55
8-
P1
A
3000 Micro GC Guides
3
The Instrument Setup window will display. It will display parameters that can be edited.
NOTE: The Instrument Setup window is configuration dependent.
Populate sample acquisition parameters as desired, such as Injector Temperature, Column
Temperature, Sample Pump time, Column Pressure, etc. Ensure that the Run Time is long
enough for the last expected peak to elute. If the elution time is unknown, conduct a run and
verify the presence of the last expected peak. If necessary, increase the Run Time.
The default Inject Time should be set between:
0-50 ms for a fixed volume injector
0-100 ms for a variable volume injector
0-350 ms for a variable large volume injector (LVI)
0-250 ms for a backflush injector
Table 7-1 Creating a method (continued)
Step
Description