The Instrument Setup window will display. It will display parameters that can be edited.

NOTE: The Instrument Setup window is configuration dependent.

Populate sample acquisition parameters as desired, such as Injector Temperature, Column
Temperature
, Sample Pump time, Column Pressure, etc. Ensure that the Run Time is long
enough for the last expected peak to elute. If the elution time is unknown, conduct a run and
verify the presence of the last expected peak. If necessary, increase the Run Time.

The default Inject Time should be set between:

0-50 ms for a fixed volume injector

0-100 ms for a variable volume injector

0-350 ms for a variable large volume injector (LVI)

0-250 ms for a backflush injector

Table 7-1 Creating a method (continued)

Step

Description