Do not turn off power. Allow three minutes for the instrument to shut down.
A The system is shutting down... message will display.

CAUTION

Do not turn off the power without waiting at least three minutes.
3000 Micro GC must update configuration files before power is
turned off. If power is turned off too soon, the files could be
corrupted, making 3000 Micro GC unusable.

7

Set the front panel power switch to the Off position.

8

Remove the two screws on each side of the top cover using a Torx T-20 or Phillips Head
screwdriver. Lift and remove the top cover.

WARNING - Risk Of Electric Shock

Electrostatic Discharge can damage electronic components. Wear
a grounded wrist strap to avoid damaging 3000 Micro GC.

Table 18-1 Changing a standard module (continued)

Step

Description