After selecting the address, click Restart.

33

A The system is shutting down... message is displayed.

Allow three minutes for the instrument to update configuration files.

CAUTION

Do not turn off the power without waiting at least three minutes.
3000 Micro GC must update configuration files before power is
turned off. If power is turned off too soon, the files could be
corrupted and render 3000 Micro GC unusable.

34

Open the web browser and type the 3000 Micro GC IP Address in the address bar. Verify that
the recently installed replacement module is available in the module list.

35

Once the replacement standard module is installed, update EZ IQ with the revised
3000 Micro GC hardware configuration. For detailed instructions on how to configure EZ IQ,
see 074-537 EZ IQ Installation Guide.

Table 18-1 Changing a standard module (continued)