4 - 2

PN

07

4-

55

8-

P1

A

3000 Micro GC Guides

6

Under Connection, select the carrier gas connection that corresponds to the desired GC
module. Click Submit.

NOTE: Entries m1 and m2 represent channels A and B, respectively.

Entries m3 and m4 represent channels C and D, if installed.

NOTE: In the example shown above, carrier1 is assigned to modules m1 and m2.

Carrier2 is unused.

7

Select the new carrier gas type from the drop-down list. Click Submit.

8

The column heaters will turn off to avoid thermal shock and degradation. Click Check Status to
view the current column temperatures.

Table 4-1 Changing the carrier gas configuration (continued)

Step

Description