P1
A
3000 Micro GC Guides
9
Loosen the thumbscrews in the manifold cover plate.
10
Remove the manifold cover plate. In the two screw design, slide the manifold cover plate
towards the GC modules to disengage the hook in the cover plate from the tab in the chassis.
11
Carefully remove the manifold insulation and save it for reuse.
NOTE: In the Portable 3000 Micro GC, the DC-DC converter assembly blocks access to
the GC modules. Remove the screws that secure the DC-DC converter assembly to
the side of the instrument. Without disconnecting any wires, slide the assembly off
of the standoffs and gently lift it away from the GC modules.
Table 18-1 Changing a standard module (continued)
Step
Description
2-channel 3000 Micro GC shown
Others are similar