A

3000 Micro GC Guides

12

Use a T-20 screwdriver (PN 5182-3468) to loosen the screw in the inlet manifold fitting
at the back of the GC module.

13

Use a 1/4 in. open-ended wrench to diconnect the inlet manifold fitting from the GC module
input fitting.

14

Disconnect the communication cables leading from the GC modules (if connected) from the GC
module communications board. Disconnect the power cables leading to the fan (if installed).

Table 18-1 Changing a standard module (continued)

Step

Description

Loosen screw

Manifold fitting screw

2-channel 3000 Micro GC shown

Others are similar

GC module

Inlet

Disconnect

inlet fittings

2-channel 3000 Micro GC shown

Others are similar

manifold

Back of GC

Remove cable

Remove
cables

Communication

board

Back view of GC modules

module