Use a T-20 screwdriver (PN 5182-3468) to loosen the screw in the inlet manifold fitting
at the back of the GC module.
13
Use a 1/4 in. open-ended wrench to diconnect the inlet manifold fitting from the GC module
input fitting.
14
Disconnect the communication cables leading from the GC modules (if connected) from the GC
module communications board. Disconnect the power cables leading to the fan (if installed).
Table 18-1 Changing a standard module (continued)
Step
Description
Loosen screw
Manifold fitting screw
2-channel 3000 Micro GC shown
Others are similar
GC module
Inlet
Disconnect
inlet fittings
2-channel 3000 Micro GC shown
Others are similar
manifold
Back of GC
Remove cable
Remove
cables
Communication
board
Back view of GC modules
module