15

Re-connect the communications cables.

NOTE: Do not connect more than two GC modules in series per communications board

connection.

NOTE: The GC modules and communications board use parallel communications;

both connectors on each item function similarly.

16

Connect the inlet manifold to the GC module input fitting and tighten it finger tight.
Use a 5/16 in. open-ended wrench, to tighten 1/4-turn past finger tight.

17

Tighten the screw in the mounting flange.

18

Carefully replace the inlet manifold insulation.

19

Install the manifold cover plate and the top cover.

20

Turn the 3000 Micro GC power switch, located on its front, to the on position. The switch will
illuminate green to indicate that 3000 Micro GC is on.

Wait three minutes for the instrument to load.

21

Open the web browser and type the 3000 Micro GC IP Address in the address bar. Verify that
the newly installed replacement module is available in the module list.

22

Once the replacement GC module is installed, update EZ IQ with the updated 3000 Micro GC
hardware configuration. For detailed instructions on how to configure EZ IQ, consult the
PN 074-537-P1A EZ IQ Installation Guide.