Caution – INFICON 3000 Micro GC Gas Analyzer Reference Guide User Manual

Page 25

Advertising
background image

4 - 3

PN

07

4-

55

8-

P1

A

3000 Micro GC Guides

9

When the column temperatures are below 60ºC, click Continue.

NOTE: Closing the browser or clicking back will cancel any changes.

10

Change the carrier gas supply connections at the instrument back panel, and check for leaks.

HINT: Label the new carrier gas type at the back of the instrument, for quick visual reference

in the future.

11

A window will prompt to restart the instrument:

Click Restart. Allow three minutes for the instrument to update configuration files.

CAUTION

Do not turn off the power without waiting at least three minutes.
3000 Micro GC must update configuration files before power is
turned off. If power is turned off too soon, the files could be
corrupted and render 3000 Micro GC unusable.

12

To run samples using the reconfigured gas type, click (Windows) Start >> Programs >>
Chromatography >>
EZ IQ Config. Update 3000 Micro GC configuration in EZ IQ.
See

3000 Micro GC Guide 6, Configuring a 3000 Micro GC with EZ IQ

.

13

A new EZ IQ method must now be created for the instrument.

Table 4-1 Changing the carrier gas configuration (continued)

Step

Description