19 - 2

PN

07

4-

55

8-

P1

A

3000 Micro GC Guides

4

Carefully remove the manifold insulation. Save for reuse.

NOTE: In the Portable 3000 Micro GC, the DC-DC converter assembly blocks access to

the GC modules. Remove the screws that secure the DC-DC converter assembly to
the side of the instrument. Without disconnecting any wires, slide the assembly off
of the standoffs and gently lift it away from the GC modules.

5

Use a T-20 screwdriver to loosen the screw in the inlet manifold fitting at the back of the
GC module.

6

Use a 1/4 in. open-ended wrench to disconnect the inlet manifold fitting from the GC module
input fitting.

Table 19-1 Changing a Performance Enhanced module (continued)

Step

Description

2-channel 3000 Micro GC shown

Others are similar

Loosen screw

Manifold fitting screw

2-channel 3000 Micro GC shown

GC module

Inlet

Disconnect

inlet fittings

2-channel 3000 Micro GC shown
Others are similar

manifold