Connecting to 3000 Micro GC

via the Static IP Address

Table 2-1 Connecting to the instrument via the static IP address

Step

Description

1

3000 Micro GC has both a static IP address and a dynamic IP address. The factory default
dynamic IP address is set at 10.1.1.101. In the event that the IP address is modified or lost, it is
possible to access the instrument through its static IP address: 192.168.1.99.

NOTE: It is necessary to access the IP address via a direct crossover cable, in order to avoid

potential routing issue in a local area network (LAN).

2

Connect one end of the crossover cable (PN 5183-4649 provided in the ship kit) to a
TCP/IP ethernet port on the PC. Connect the other end to the ethernet port on the back of
3000 Micro GC.

3

The 3000 Micro GC power switch is located on its front panel. Turn the switch to the on position.
The switch will illuminate green to indicate that 3000 Micro GC is on. Wait three minutes for the
instrument to load.

4

Click (Windows) Start >> Connect to >> Show All Connections.

NOTE: Navigation path my vary, depending on what version Windows is running.

5

On the Network Connections window, double-click Local Area Connection.