Tilt the back of the GC module up until the mounting flange clears the alignment pins. Slide the
GC module towards the back of the instrument until it clears the inlet manifold frame and can
be lifted and removed from the front.

16

Inspect the gang block fitting on the bottom of the chassis to make sure the mating surface
is free of dust and particles.

17

On the replacement module, remove the small metal plate covering the O-rings on the upper
gang block.

Table 18-1 Changing a standard module (continued)

Step

Description

Tilt GC module

Mounting flange

Mounting flange

Tilt until mounting

flange is free from

aligning pins

2-channel 3000 Micro GC shown
Others are similar

Gang block, lower half

Alignment pins

Metal plate covering

O-rings and upper

gang block

Retaining screw