When finished, click Start.

15

When the analysis is complete, the chromatogram will be integrated, and the areas for the
peaks identified as calibration compounds will be entered into the method.

16

With the calibration gas still connected and flowing, repeat steps 4 through 14 for each replicate.
Ensure that each Sample ID is unique, for example calgas1rep2, calgas1rep3, etc.

17

After completing a calibration in the window, a calibration curve and associated data can be
reviewed by clicking Method >> Review Calibration. Each peak will have its own calibration
curve. If desired, right click on a black area on the window and select commands such as View
Fit Type
, Change Fit Type, and a Concentration Calculator.

The peak Area will be reported as the average area of the replicates.