Guide 18 changing a standard module, 3000 micro gc guide 18 – INFICON 3000 Micro GC Gas Analyzer Reference Guide User Manual

Page 67

Advertising
background image

18 - 1

PN

07

4-

55

8-

P1

A

3000 Micro GC Guides

3000 Micro GC Guide 18

Changing a Standard Module

Table 18-1 Changing a standard module

Step

Description

1

A Standard GC module must be decommissioned before it can be removed. 3000 Micro GC
internally communicates to each installed GC module via a unique module address.

For a 1-2 channel or portable instrument, the module addresses are 1 and 2.

For a 3-4 channel instrument, the module addresses are 1, 2, 3 and 4.

Decommissioning a Standard GC module will disable the GC module’s address. Open the web
browser and type the 3000 Micro GC IP Address in the address bar; press Enter to display the
embedded web page.

NOTE: At the factory, the instrument IP address is set to 10.1.1.101.

2

Click the Utilities tab.

Click Change Module Configuration.

NOTE: Two caution messages will display. Click OK on each to continue.