Replacing a Standard Module with a

Performance Enhanced Module

Table 20-1 Changing a standard module

Step

Description

1

A Standard GC module must be decommissioned before it can be removed. 3000 Micro GC
internally communicates to each installed GC module using a unique module address.

For a 1-2 channel or portable instrument, the module addresses are 1 and 2.
For a 3-4 channel instrument, the module addresses are 1, 2, 3 and 4.

Decommissioning a Standard GC module will disable the GC module’s address.

Open the web browser and type the 3000 Micro GC IP Address in the address bar, then press
Enter to display the embedded web page.

NOTE: At the factory, the instrument IP address is set to 10.1.1.101.

2

Click the Utilities tab.

Click Change Module Configuration.

3

Two caution messages will display. Click OK on each to continue.