Guide 19 changing a performance enhanced module, 3000 micro gc guide 19, Warning - risk of electric shock – INFICON 3000 Micro GC Gas Analyzer Reference Guide User Manual

Page 79

Advertising
background image

19 - 1

PN

07

4-

55

8-

P1

A

3000 Micro GC Guides

3000 Micro GC Guide 19

Changing a Performance Enhanced Module

Table 19-1 Changing a Performance Enhanced module

Step

Description

1

To replace a Performance Enhanced module with another Performance Enhanced module, turn
of the instrument by setting the front panel power switch to the Off position. Remove the two
screws on each side of the top cover using a Torx T-20 or Phillips Head screwdriver. Lift and
remove the top cover.

WARNING - Risk Of Electric Shock

Electrostatic Discharge can damage electronic components. Wear
a grounded wrist strap to avoid damaging 3000 Micro GC.

2

Loosen the thumbscrews in the manifold cover plate.

3

Remove the manifold cover plate. Slide the manifold cover plate towards the GC module to
disengage the hook in the cover plate from the tab in the chassis.

Advertising