3.4 System Menu

The System Menu sets values that pertain to the overall functions of the SQM-160
and to the vacuum system's setup. System Menu parameters apply to all films.

1

Press Program to enter program mode.

2

Press Prev to enter the System Menu.

3

Use Next and Prev to move through the system parameters.

4

Use the Control Knob to adjust the parameter value shown in Display 2 to the
desired setting.

5

Press the Control Knob or Next to save the displayed value and move to the
next material parameter. Press Clear to abandon the change and return to the
original setting.

6

Press Program to exit the System Menu and return to normal mode.

Figure 3-3 System Parameters Menu

Program

Prev

- TIMEBASE
- SIM MODE
- DISPLAY

- RATE RES
- RATEFILT
- RELAY 2
- SAMPLING
- BAUDRATE
- ETCH
- xTOOLING
- RELAYS
- FM IN/MAX
- R/T BNDS

- THCK (A)
- THCK(Nm)
- MASS
- FREQ

- TIME SP
- DUAL
- SENS 2

- FREQ MIN
- FREQ MAX

- xTLTOOL1

- xTLTOOL6

- RELAY 1

- RELAY 4

System Parameters Menu

Sub-Menu

- RATE MIN
- RATE MAX
- THICKMIN
- THICKMAX

Table 3-2 System Menu

Display

Description

Range

Default

Units

TIMEBASE

Time required for a measurement.
Longer times yield higher accuracy.

0.15 – 2.00

0.25

Sec.

SIM MODE

Simulates sensor inputs.

On/Off

Off

DISPLAY

Selects Rate/Thickness in Angstroms,
Rate/Thickness in Nanometers, Frequency,
or Mass (ugm/cc) display.

THCK/nÅnM/FRE

Q/MASS

Rate

RATE RES

Sets rate resolution to 0.01 or

0.1 Å/s.

Hi/Low

Low