5 command: d, 6 command: j, 7 command: l – INFICON SQM-160 Thin Film Deposition Monitor User Manual

Page 69: 8 command: m, 9 command: n, 10 command: o

Advertising
background image

5 - 9

IP

N 07

4-

51

1-

P1

C

SQM-160 Operating Manual

5.4.5 Command: D

Parameters: . . . . . . . . . . . . . . . . . . . 1 to 9

Description: . . . . . . . . . . . . . . . . . . . Sets the active film.

Example: . . . . . . . . . . . . . . . . . . . . . D1

Set the active film to Film #1

5.4.6 Command: J

Parameters: . . . . . . . . . . . . . . . . . . . None.

Description: . . . . . . . . . . . . . . . . . . . Read the number of channels installed. The

number of channels will be either an ASCII
two or six.

Example: . . . . . . . . . . . . . . . . . . . . . J A2

The SQM-160 has two channels available.

5.4.7 Command: L

Parameters: . . . . . . . . . . . . . . . . . . . [1..6]

Description: . . . . . . . . . . . . . . . . . . . Read the current Rate for a channel.

Example: . . . . . . . . . . . . . . . . . . . . . L1 A9.32

Channel one's rate is 9.32 Å/S

5.4.8 Command: M

Parameters: . . . . . . . . . . . . . . . . . . . None.

Description: . . . . . . . . . . . . . . . . . . . Read the current Average Rate.

Example: . . . . . . . . . . . . . . . . . . . . . M A10.42

Average Rate is 10.42 Å/S

5.4.9 Command: N

Parameters: . . . . . . . . . . . . . . . . . . . [1..6]

Description: . . . . . . . . . . . . . . . . . . . Read the current thickness for a channel.

Example: . . . . . . . . . . . . . . . . . . . . . N4 A1.187

Channel four's Thickness is 1.187 kÅ.

5.4.10 Command: O

Parameters: . . . . . . . . . . . . . . . . . . . None.

Description: . . . . . . . . . . . . . . . . . . . Read the current Average Thickness

Example: . . . . . . . . . . . . . . . . . . . . . O A2.376

The current Average Thickness is 2.376

kЕ.

Advertising