11 command: p, 12 command: r, 13 command: s – INFICON SQM-160 Thin Film Deposition Monitor User Manual

Page 70: 14 command: t, 15 command: u

Advertising
background image

5 - 10

IP

N 07

4-

51

1-

P1

C

SQM-160 Operating Manual

5.4.11 Command: P

Parameters: . . . . . . . . . . . . . . . . . . . [1..6]

Description: . . . . . . . . . . . . . . . . . . . Read the current Frequency for a channel.

Example: . . . . . . . . . . . . . . . . . . . . . P2 A5701563.2

Channel two's current Frequency

5701563.2 Hz

5.4.12 Command: R

Parameters: . . . . . . . . . . . . . . . . . . . [1..6]

Description: . . . . . . . . . . . . . . . . . . . Read the Crystal Life for a channel.

Example: . . . . . . . . . . . . . . . . . . . . . R3 A57.82

Channel three's remaining life is 57.82%.

5.4.13 Command: S

Parameters: . . . . . . . . . . . . . . . . . . . None.

Description: . . . . . . . . . . . . . . . . . . . Zero Average Thickness and Rate.

Example: . . . . . . . . . . . . . . . . . . . . . S A

5.4.14 Command: T

Parameters: . . . . . . . . . . . . . . . . . . . None.

Description: . . . . . . . . . . . . . . . . . . . Zero Time

Example: . . . . . . . . . . . . . . . . . . . . . T A

Zeroes time display on SQM-160.

5.4.15 Command: U

Parameters: . . . . . . . . . . . . . . . . . . . 0,1, or ?

Description: . . . . . . . . . . . . . . . . . . . Toggles shutter open/closed or reads shutter

state.

Example: . . . . . . . . . . . . . . . . . . . . . U1 A

Shutter is opened

U? A1

Shutter Status is open

U0 A

Shutter is closed

Advertising