5 - 10
IP
N 07
4-
51
1-
P1
C
SQM-160 Operating Manual
5.4.11 Command: P
Parameters: . . . . . . . . . . . . . . . . . . . [1..6]
Description: . . . . . . . . . . . . . . . . . . . Read the current Frequency for a channel.
Example: . . . . . . . . . . . . . . . . . . . . . P2 A5701563.2
Channel two's current Frequency
5701563.2 Hz
5.4.12 Command: R
Parameters: . . . . . . . . . . . . . . . . . . . [1..6]
Description: . . . . . . . . . . . . . . . . . . . Read the Crystal Life for a channel.
Example: . . . . . . . . . . . . . . . . . . . . . R3 A57.82
Channel three's remaining life is 57.82%.
5.4.13 Command: S
Parameters: . . . . . . . . . . . . . . . . . . . None.
Description: . . . . . . . . . . . . . . . . . . . Zero Average Thickness and Rate.
Example: . . . . . . . . . . . . . . . . . . . . . S A
5.4.14 Command: T
Parameters: . . . . . . . . . . . . . . . . . . . None.
Description: . . . . . . . . . . . . . . . . . . . Zero Time
Example: . . . . . . . . . . . . . . . . . . . . . T A
Zeroes time display on SQM-160.
5.4.15 Command: U
Parameters: . . . . . . . . . . . . . . . . . . . 0,1, or ?
Description: . . . . . . . . . . . . . . . . . . . Toggles shutter open/closed or reads shutter
state.
Example: . . . . . . . . . . . . . . . . . . . . . U1 A
Shutter is opened
U? A1
Shutter Status is open
U0 A
Shutter is closed