Limited warranty, Warning – SATEC EM920 Operation Manual User Manual

Page 2

Advertising
background image

2

EM920 Power Quality and Revenue Meter

LIMITED WARRANTY

The manufacturer offers the customer a 24-month functional warranty on the
instrument for faulty workmanship or parts from date of dispatch from the
distributor. In all cases, this warranty is valid for 36 months from the date of
production. This warranty is on a return to factory basis.

The manufacturer does not accept liability for any damage caused by
instrument malfunction. The manufacturer accepts no responsibility for the
suitability of the instrument to the application for which it was purchased.

Failure to install, set up or operate the instrument according to the instructions
herein will void the warranty.

Only a duly authorized representative of the manufacturer may open your
instrument. The unit should only be opened in a fully anti-static environment.
Failure to do so may damage the electronic components and will void the
warranty.

The greatest care has been taken to manufacture and calibrate your
instrument. However, these instructions do not cover all possible contingencies
that may arise during installation, operation or maintenance, and all details
and variations of this equipment are not covered by these instructions.

For additional information regarding installation, operation or maintenance of
this instrument, contact the manufacturer or your local representative or
distributor.

WARNING

Read the instructions in this manual before performing installation,
and take note of the following precautions:

Ensure that all incoming AC power and other power sources
are turned OFF before performing any work on the instrument.
Failure to do so may result in serious or even fatal injury
and/or equipment damage.

Before connecting the instrument to the power source, check
the labels on the back of the instrument to ensure that your
instrument is equipped with the appropriate power supply
voltage, input voltages and currents.

Under no circumstances should the instrument be
connected to a power source if it is damaged.

To prevent potential fire or shock hazard, do not expose the
instrument to rain or moisture.

The secondary of an external current transformer must never
be allowed to be open circuit when the primary is energized.
An open circuit can cause high voltages, possibly resulting in
equipment damage, fire and even serious or fatal injury.
Ensure that the current transformer wiring is secured using an
external strain relief to reduce mechanical strain on the screw
terminals, if necessary.

Only qualified personnel familiar with the instrument and its
associated electrical equipment must perform setup
procedures.

Do not open the instrument under any circumstances when it is
connected to a power source.

Advertising