3 properties – swing head, Properties – swing head – Metrohm 900 Touch Control Manual User Manual
Page 110

11.5 USB Sample Processor
■■■■■■■■■■■■■■■■■■■■■■
94
■■■■■■■■
900 Touch Control
off
No check takes place.
Rotation increment
Amount by which the sample rack can be rotated forward or backward
relative to its current position. This parameter is required in the command
MOVE for the settings Destination = rotate + and rotate -.
Input range
0 - 353.99 °
Default value
5.00 °
Axial distance
Distance between the axis of rotation of the sample rack and swing axis of
the robotic arm.
for 814 USB Sample Processor XL:
Input range
100.0 - 300.0 mm
Default value
166.0 mm
for 815 Robotic USB Sample Processor XL, 855 Robotic Titrosampler:
Input range
100.0 - 300.0 mm
Default value
196.0 mm
11.5.3
Properties – Swing Head
List of devices: Sample Processor
▶ Edit ▶ Swing Head ▶ Properties
The dialog Edit device / Properties shows a list of all external positions
with the assigned swing angle and the specific work position for each.
[Ext. pos. 1-4]
Edit the settings that apply for all external positions (see "Properties –
External positions 1-4", page 95).