28 Manual control
900 Touch Control
■■■■■■■■
257
With the function [Dosing], dosing is continuous for as long as you hold
the button [Dose] pressed down.
Proceed as follows:
1 Configure the dosing function
NOTE
■
The dosing and filling rates should be decreased for viscous liq-
uids.
■
The maximum dosing and filling rates depend on the cylinder
volume (see Chapter 32, page 488).
■
Enter the dosing rate.
■
Enter the filling rate.
2 Start dosing
■
Tap on [Dose].
The volume dosed is displayed. When the volume of one cylinder has
been dosed, the dosing cylinder will be refilled automatically.
3 Fill the cylinder
■
Tap on [Fill].
The dosing cylinder is filled. The displayed volume value will be reset
to 0.0000 mL.
NOTE
If you exit the dialog with the fixed key [
], the dosing cylinder
will be filled automatically.
Parameter description
Dosing rate
Rate at which it is dosed.
Input range
0.01 - 166.00 mL/min
Selection
maximum | dynamic
Default value
maximum
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 32, page 488).