8.4 Dosing unit
■■■■■■■■■■■■■■■■■■■■■■
50
■■■■■■■■
900 Touch Control
Diameter
Diameter of the tubing.
Input range
0.0 - 9.9 mm
Default value
2.0 mm
Fill port
Port
Port to be used as fill port for the PREP and EMPTY functions (see Figure
8, page 51).
Selection
Port 1 | Port 2 | Port 3 | Port 4
Default value
Port 2
Length
Length of the tubing.
Input range
0.0 - 999.9 cm
Default value
25.0 cm
The setting 0.0 means that this tubing will neither
be rinsed nor emptied.
Diameter
Diameter of the tubing.
Input range
0.0 - 9.9 mm
Default value
2.0 mm
Special port
Port
Port to be used as special port for the PREP and EMPTY functions (see
Figure 8, page 51).
Selection
Port 1 | Port 2 | Port 3 | Port 4
Default value
Port 4
Length
Length of the tubing.
Input range
0.0 - 999.9 cm
Default value
0.0 cm
The setting 0.0 means that this tubing will neither
be rinsed nor emptied.
Diameter
Diameter of the tubing.