3 titration parameters, Titration parameters – Metrohm 900 Touch Control Manual User Manual

Page 372

Advertising
background image

29.7 STAT titrations (STAT)

■■■■■■■■■■■■■■■■■■■■■■

356

■■■■■■■■

900 Touch Control

Min. rate in µL/min = 0.1 · expected reaction rate in µL/min

Input range

0.01 - 9999.00 µL/min

Default value

10.00 µL/min

Table 12

Default values of the predefined titration rates for STAT

Reaction rate

50 µL/min

100 µL/min

500 µL/min

Dynamics

– pH

– U

1.000

60.0 mV

1.000

60.0 mV

1.000

60.0 mV

Max. rate

0.25 mL/min

0.75 mL/min

2.00 mL/min

Min. rate

5.00 µL/min

10.00 µL/min

40.00 µL/min

Dosing and control range
Outside the control range the dosing rate is mainly characterized by the
maximum rate. The parameters Max. rate and Dynamics should be opti-
mized together in such a way that the titration does not overshoot too
much when the control point is reached. The control range should be
selected in such a way that the measured value remains within this range
when dosing stops. A slightly larger control range should be defined for
slow reactions (e.g. pH = 3, U = 180 mV). The setpoint value is often set
with a SET pretitration before the substrate is added. This means that the
STAT titration can be started with a small deviation from the control point.

Within the control range the dosing rate is mainly characterized by the
minimum rate.

29.7.3

Titration parameters

Under [Titration parameters], the parameters influencing the sequence
of the entire titration are defined.

Titration direction

The titration direction is normally determined automatically from the start
measured value and the set control point. It is recommended that, when-
ever possible, a positive or negative modification of the measured value is
entered.

Selection

+ | - | auto

Default value

auto

+
Positive measured value modification, i.e. in the direction of a higher
pH value or greater voltage.