16 Methods

900 Touch Control

■■■■■■■■

177

■

Enter the value 10 mL as Sample size.

3 Display the main dialog

Tap on the fixed key [

].

The data is saved and the adjusted main dialog appears.

Only Identification 1 is shown with the title "Batch." The fixed sample
size "10 mL" is shown, but cannot be edited.

Defining limit values for the sample size
In the dialog Sample data / Sample size limits, you can define the limit
values for the sample size. These values are not monitored during sample
data input, but rather:

■

when the determination is started.

■

when the determination is finished.

■

upon automatic sample data query via the command REQUEST.

■

when the determination is being recalculated.

An entry is made in the message list automatically when these limits are
infringed (see Chapter 24.2, page 216).

Monitoring sample size limits

on | off (Default value: off)

If this parameter is activated, then the sample size will be monitored.

Lower limit

Input range

–999999999 - 9999999999

Default value

0