16 Methods
900 Touch Control
■■■■■■■■
177
■
Enter the value 10 mL as Sample size.
3 Display the main dialog
Tap on the fixed key [
].
The data is saved and the adjusted main dialog appears.
Only Identification 1 is shown with the title "Batch." The fixed sample
size "10 mL" is shown, but cannot be edited.
Defining limit values for the sample size
In the dialog Sample data / Sample size limits, you can define the limit
values for the sample size. These values are not monitored during sample
data input, but rather:
■
when the determination is started.
■
when the determination is finished.
■
upon automatic sample data query via the command REQUEST.
■
when the determination is being recalculated.
An entry is made in the message list automatically when these limits are
infringed (see Chapter 24.2, page 216).
Monitoring sample size limits
on | off (Default value: off)
If this parameter is activated, then the sample size will be monitored.
Lower limit
Input range
–999999999 - 9999999999
Default value
0