Metrohm 900 Touch Control Manual User Manual
Page 275

■■■■■■■■■■■■■■■■■■■■■■
28 Manual control
900 Touch Control
■■■■■■■■
259
■
Enter the dosing rate.
■
Enter the filling rate.
2 Start dosing
■
Tap on [Start].
The volume display is shown. When the volume of one cylinder has
been dosed, the dosing cylinder will be refilled automatically.
Continuous dosing is paused with [Hold]. The label changes to
[Continue]. This can be used to continue the run.
3 Fill the cylinder
With the default settings (see "Fill automatic.", page 260) the dos-
ing cylinder is filled automatically.
Otherwise:
■
Tap on [Fill].
The dosing cylinder is filled. The displayed volume value will be reset
to 0.0000 mL.
Parameter description
Volume
Volume which is dosed.
Input range
0.00000 - 99999.9 mL
Default value
0.10000 mL
Dosing rate
Rate at which it is dosed.
Input range
0.01 - 166.00 mL/min