2 standard addition, Standard addition – Metrohm 900 Touch Control Manual User Manual

Page 397

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

29 Parameters

900 Touch Control

■■■■■■■■

381

Selection

constant | ISO / DIN

constant
The temperature compensation is carried out with the value entered
for the temperature coefficient.
ISO / DIN
For ground water, spring water and surface water, the temperature
coefficients according to standard ISO 7888:1985 (German version:
DIN EN 27888:1993) are stored in the system.

Temp. coefficient

This parameter can only be edited with Temp. compensation = con-
stant
.

Temperature coefficient belonging to the reference temperature entered
above. With this temperature coefficient, the conductivity is converted to
the reference temperature.

Input range

0.00 - 9.99 %/°C

Default value

0.00 %/°C

29.9.2

Standard addition

Under [Standard addition], the parameters for the concentration mea-
surement by means of standard addition are defined. This button is avail-
able only with the MEAS Conc command (concentration measurement).

Addition

Selection of the addition mode.

Selection

off | auto | auto dos | manual

Default value

off

off
The concentration measurement is carried out without standard addi-
tion, i.e. as a direct measurement.
auto
The standard solution is dosed automatically. The individual volume
steps are calculated in the way that a certain voltage difference is
reached. This voltage difference (Delta U) is defined under Standard
addition

▶ Dosing device.

auto dos
The standard solution is dosed automatically. The individual volume
steps are defined under Standard addition

▶ Addition increments.

manual
The standard solution is dosed manually. The individual volume steps
are defined under Standard addition

▶ Addition increments.