1 carrying out a software update – Metrohm 900 Touch Control Manual User Manual
Page 524

32.9 Diagnosis
■■■■■■■■■■■■■■■■■■■■■■
508
■■■■■■■■
900 Touch Control
They can be recognized by means of the two-digit language code in the
file name. The file name has the following structure:
■
5XXXyZZZML.BIN
5XXXyZZZML.MBIN
– XXX = instrument type (i.e. "900" for the 900 Touch Control)
– y = version number of the language package
– ZZZ = program version
32.9.7.1
Carrying out a software update
CAUTION
Make sure that the power supply is guaranteed during the entire
update process. Otherwise there is the chance that the instrument will
no longer be able to be switched on and that it will need to be sent in
for repair.
Proceed as follows:
1 Copying files to an external storage medium
NOTE
In the case of the program versions and language files for the
Touch Control, it is imperative that both the BIN file and the asso-
ciated MBIN file be copied.
■
Copy the files to the "SwUpdates" directory.