29.9 Measurements (MEAS)

■■■■■■■■■■■■■■■■■■■■■■

382

■■■■■■■■

900 Touch Control

Concentration standard

Concentration of the standard solution used. When Addition = manual
is defined, a numerical value has to be entered here, the selection from
list
creates an error message when starting the determination.

Input range

0.00001 - 99999.9

Selection

from list

Default value

from list

from list
The standard solution selected under Standard addition

▶ Dosing

device is used.

Unit of the standard solution used.

Entry

max. 24 characters

Selection

mol/L | mmol/L | ppm | % | g/L | mg/L | µg/L

Default value

ppm

Number of additions

Number of standard additions. The more addition steps you carry out the
more accurate the evaluation becomes.

Input range

1 - 19

Default value

3

Stop volume

Maximum permissible volume for the volume steps. If the sum of the vol-
ume steps is higher than this volume, an error message is displayed when
starting the determination.

Input range

0.00 - 9999.90 mL

Default value

99.99 mL

Volume aux. solution

Volume of the buffer solution etc. which has to be added to the sample
before the measurement can be started.

Input range

0.00 - 9999.90 mL

Default value

0.00 mL

Stirrer control
The parameters for the stirrer control during the standard addition are
edited under [Stirrer control]. Ensure that a stirrer has been selected in
the Edit command / Stirrer dialog.