29.7 STAT titrations (STAT)
■■■■■■■■■■■■■■■■■■■■■■
362
■■■■■■■■
900 Touch Control
Measuring mode pH:
Input range
0.000 - 20.000
Default value
0.020
Measuring mode U:
Input range
0.0 - 2,000.0 mV
Default value
2.0 mV
Dosing rate:
Input range
0.00 - 166.00 mL/min
Default value
0.20 mL/min
Temperature:
Input range
0.0 - 150.0 °C
Default value
0.2 °C
Action
Selection of the action which is carried out when the limits of the moni-
tored quantity are infringed.
Selection
None | Exit method | Cancel command |
Pause | Wait
None
No action will be taken.
Exit method
The method is stopped immediately.
Cancel command
The current command is stopped und the next command in the
method will be started.
Pause
Adding reagent is stopped until the dosing is continued manually with
[Continue]. If several quantities are monitored, the other quantities
will still be monitored in the background.
Wait
Adding reagent is stopped until the value of the monitored quantity is
within the limits again (incl. hysteresis). The dosing is continued auto-
matically. If several quantities are monitored, the other quantities will
still be monitored in the background.