For buret units with integrated data chip, a check is made in the method
run to verify whether the correct titrant has been attached and whether
the type of dosing drive matches. For buret units without integrated data
chip, the cylinder volume and the type of dosing drive are checked. For
the selected titrant, the validity of the titer, the working life of the titrant
and the time interval for the GLP test of the buret unit are checked at the
start of the determination.

Selection

Selection of configured titrants | not defined

Default value

not defined

not defined
No check takes place.

29.4.10 Stirrer

The parameters for the stirrer are edited under [Stirrer].

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB
connector will always be displayed.

Selection

1 | 2 | 3 | 4 | off

Default value

1

off
No stirrer will be used.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default
setting 8 corresponds to 1000 rpm. The formula for calculating the rota-
tional speed is specified in chapter 32.3, page 489. The optimum stirring
rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the
stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

–15 - 15

Default value

8

Switch off automatically

on | off (Default value: on)

If this parameter is activated, the stirrer will be switched off automatically
at the end of the titration, measurement, etc.