2 dosing fixed volumes, Dosing fixed volumes – Metrohm 900 Touch Control Manual User Manual
Page 274

28.3 Dosing
■■■■■■■■■■■■■■■■■■■■■■
258
■■■■■■■■
900 Touch Control
dynamic
This setting can only be selected when the dosing cylinder is filled. The
dosing is being carried out faster and faster until the maximum dosing
rate is reached (starting with 1 mL/min and then doubling the dosing
rate every 1.5 s).
Filling rate
Rate at which the dosing cylinder is filled.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum filling rate depends on the cylinder volume (see Chapter
32, page 488).
28.3.2
Dosing fixed volumes
Manual control: Dosing
▶ Dosing fixed vol.
You can dose a particular volume with the [Dosing fixed vol.] function.
Proceed as follows:
1 Configure the dosing function
NOTE
■
The dosing and filling rates should be decreased for viscous liq-
uids.
■
The maximum dosing and filling rates depend on the cylinder
volume (see Chapter 32, page 488).
■
Enter the desired volume.