29 Parameters

900 Touch Control

■■■■■■■■

349

NOTE

This parameter must be activated if commands follow which must not
run simultaneously with the conditioning.

Example: Emptying the titration cell followed by adding new working
medium.

29.6.5

Cell

The parameters for the titration cell are defined under [Cell].

Generator electrode

Type of generator electrode.

Selection

without diaphragm | with diaphragm

Default value

with diaphragm

without diaphragm
Generator electrode without diaphragm.
with diaphragm
Generator electrode with diaphragm.

Generator current

Polarization current at the generator electrode.

Selection

100 mA | 200 mA | 400 mA | auto

Default value

auto

auto
The current is adapted to the conductivity of the reagent and automat-
ically reduced near the endpoint.

Reagent monitoring

Selection of the reagent from the list of reagents. The selection depends
on the titration mode. Reagents are defined under System

▶ Reagents.

Selection

Selection of configured reagents | off

Default value

off

off
The reagent monitoring is not active.