5 limit values for the calibration data – Metrohm 900 Touch Control Manual User Manual
Page 91

■■■■■■■■■■■■■■■■■■■■■■
10 Sensors
900 Touch Control
■■■■■■■■
75
10.5
Limit values for the calibration data
Sensor: Edit
▶ Limit values
You can define the following limit values in the dialog Edit sensor /
Limit values:
■
Slope (pH and ion-selective electrodes)
■
Electrode zero point (pH and ion-selective electrodes)
■
Cell constant (conductivity measuring cells)
These values are monitored during the calibration. If these limits are
infringed, then a message will be displayed and you can decide whether
to accept the calibration data or not.
Monitoring slope
on | off (Default value: off)
If this parameter is activated, then the slope will be monitored.
Lower limit
pH electrodes:
Input range
–999.9 - 999.9 %
Default value
96.0 %
ISE electrodes:
Input range
–999.9 - 999.9 mV
The default value depends on the charge of the ion:
–25.0 mV (–2), –55.0 mV (–1), 55.0 mV (+1), 25.0
mV (+2).