10.5 Limit values for the calibration data
■■■■■■■■■■■■■■■■■■■■■■
76
■■■■■■■■
900 Touch Control
Upper limit
pH electrodes:
Input range
–999.9 - 999.9 %
Default value
101.0 %
ISE electrodes:
Input range
–999.9 - 999.9 mV
The default value depends on the charge of the ion:
–31.0 mV (–2), –61.0 mV (–1), 61.0 mV (+1), 31.0
mV (+2).
Monitoring pH(0)
on | off (Default value: off)
This parameter is only available with pH electrodes.
If this parameter is activated, then the electrode zero point pH(0) will be
monitored.
Lower limit
Input range
–20.000 - 20.000
Default value
6.750
Upper limit
Input range
–20.000 - 20.000
Default value
7.250
Monitoring E(0)
on | off (Default value: off)
This parameter is only available with ISE electrodes.
If this parameter is activated, then the electrode zero point E(0) will be
monitored.
Lower limit
Input range
–2000.0 - 2000.0 mV
Default value
–2000.0 mV
Upper limit
Input range
–2000.0 - 2000.0 mV
Default value
2000.0 mV
Monitoring cell constant
on | off (Default value: off)
This parameter is only available for conductivity measuring cells.