Figure 18, Action "wait – Metrohm 900 Touch Control Manual User Manual
Page 380

29.7 STAT titrations (STAT)
■■■■■■■■■■■■■■■■■■■■■■
364
■■■■■■■■
900 Touch Control
Upper hysteresis
Lower hysteresis
Upper limit
Lower limit
Wait
Monitored
quantity
Time
Figure 18
Action "Wait"
Communication
Under [Communication], remote signals or RS-232 commands can be
assigned to the monitored quantities. These remote signals or RS-232
commands are always sent, independent of the action defined for a limit
infringement (see "Action", page 362).
[New]
Define a new assignment.
[Delete]
Delete the selected assignment.
[Edit]
Edit the settings for the selected assignment.
Monitored quantity
Selection of the monitored quantity to which a remote signal or an
RS-232 command is to be assigned.
Selection
Measured value | Temperature | Dosing rate |
any
any
If several quantities are monitored, the remote signal or RS-232 com-
mand set is sent as soon as a limit of one of the monitored quantities is
exceeded.
Infringed limit
Indicating the limit whose infringement causes the remote signal or
RS-232 command defined in the following to be sent.
Selection
upper | lower | any | back in range