29.8 Manual titrations (MAT)

■■■■■■■■■■■■■■■■■■■■■■

376

■■■■■■■■

900 Touch Control

The dosing rate must not exceed 75 % of the filling rate of the smaller
cylinder. These values are listed in the following table, valid at maxi-
mum filling rate:

Table 14

Maximum dosing rate for different dosing cylinders

Cylinder volume

maximum dosing rate

Exchange unit

Dosing unit

1 mL

2.25 mL/min

2 mL

5.00 mL/min

5 mL

11.25 mL/min

12.50 mL/min

10 mL

22.50 mL/min

25.00 mL/min

20 mL

45.00 mL/min

50.00 mL/min

50 mL

112.50 mL/min

124.50 mL/min

29.8.5

Stirrer

The parameters for the stirrer are edited under [Stirrer].

Stirrer

Selection of the MSB connector the stirrer is connected to. Every MSB
connector will always be displayed.

Selection

1 | 2 | 3 | 4 | off

Default value

1

off
No stirrer will be used.

Stirring rate

Setting the stirring rate. It can be set in steps of –15 to +15. The default
setting 8 corresponds to 1000 rpm. The formula for calculating the rota-
tional speed is specified in chapter 32.3, page 489. The optimum stirring
rate can be tested in the manual control.

The algebraic sign of the stirring rate changes the direction in which the
stirring is done. When the stirrer is viewed from above, this means:

"+": counterclockwise rotation

"–": clockwise rotation

Input range

–15 - 15

Default value

8