3 stop conditions, Stop conditions – Metrohm 900 Touch Control Manual User Manual

Page 315

Advertising
background image

■■■■■■■■■■■■■■■■■■■■■■

29 Parameters

900 Touch Control

■■■■■■■■

299

Titration rate

slow

optimal

fast

Signal drift

– pH, U and Ipol

– Upol

20.0 mV/min

20.0 µA/min

50.0 mV/min

50.0 µA/min

80.0 mV/min

80.0 µA/min

Min. waiting time

0 s

0 s

0 s

Max. waiting time

38 s

26 s

21 s

29.2.3

Stop conditions

The conditions for canceling the titration are defined under [Stop condi-
tions]
.

Stop volume

The titration is stopped when the specified volume has been dosed since
the start of the titration. This volume should be adjusted to the size of the
titration vessel in order to prevent the contents from running over.

Input range

0.00000 - 9999.99 mL

Default value

100,000 mL

Selection

off

Stop meas. value

The titration is canceled when the specified measured value has been
reached since the start of the titration.

The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.

Measuring mode pH:
Input range

–20.000 - 20.000

Selection

off

Default value

off

Measuring mode U, Ipol:
Input range

–2,000.0 - 2,000.0 mV

Selection

off

Default value

off

Measuring mode Upol:
Input range

–200.0 - 200.0 µA

Selection

off

Default value

off

Advertising