28.2 Measuring

■■■■■■■■■■■■■■■■■■■■■■

252

■■■■■■■■

900 Touch Control

Measuring mode

Selection of the measuring mode. Only those measuring modes are dis-
played which are advisable for the selected sensor.

Selection

pH | U | Ipol | Upol | Conc | Cond | T

pH
potentiometric pH measurement
U
potentiometric voltage measurement
Ipol
voltametric measurement with selectable polarization current
Upol
amperometric measurement with selectable polarization voltage
Conc
Concentration measurement as direct measurement
Cond
Conductivity measurement
T
Temperature measurement

Temperature

This parameter is not available for conductivity measuring cells.

Temperature entered manually. If a temperature sensor is connected then
the temperature will be measured continuously. This value is used for tem-
perature correction in pH and concentration measurements.

Input range

–20.0 - 150.0 °C

Default value

25.0 °C

Measuring temp.

This parameter is available only with conductivity measurements.

Temperature entered manually. If a temperature sensor is connected then
the temperature will be measured continuously.

Input range

–20.0 - 150.0 °C

Default value

25.0 °C

[Temp. compens.]

This button is available only with conductivity measurements.

Edit the parameters for converting the measured conductivity to a refer-
ence temperature.

[Info sensor]

Display information on the sensor.