29.2 Monotonic equivalence point titrations (MET)
■■■■■■■■■■■■■■■■■■■■■■
294
■■■■■■■■
900 Touch Control
Measuring mode pH:
Input range
–20.000 - 20.000
Selection
off
Default value
off
Measuring mode U, Ipol:
Input range
–2,000.0 - 2,000.0 mV
Selection
off
Default value
off
Measuring mode Upol:
Input range
–200.0 - 200.0 µA
Selection
off
Default value
off
Start slope
Before starting the titration, it will be dosed until this slope (measured
value per volume) is reached. If the start slope is reached by the dosing of
a start volume, then the titration starts directly.
Measuring mode pH:
Input range
00.000 - 9,999 pH/mL
Selection
off
Default value
off
Measuring mode U, Ipol:
Input range
0 - 999 mV/mL
Selection
off
Default value
off
Measuring mode Upol:
Input range
0 - 99 µA/mL
Selection
off
Default value
off
Dosing rate
Rate to be dosed with until the start measured value or the start slope is
reached.
Input range
0.01 - 166.00 mL/min
Default value
5.00 mL/min
Selection
maximum
maximum
The maximum dosing rate depends on the cylinder volume (see Chap-
ter 32, page 488).