28.5 Conditioning a titration cell

■■■■■■■■■■■■■■■■■■■■■■

264

■■■■■■■■

900 Touch Control

The algebraic sign of the stirring rate changes the direction in which the
stirring is done. When the stirrer is viewed from above, this means:

■

"+": counterclockwise rotation

■

"–": clockwise rotation

Input range

–15 - 15

Default value

8

NOTE

If a stirrer is used in the determination run, it can nevertheless be con-
trolled manually at the same time. For example, you can use the man-
ual control to reduce the stirring rate of a stirrer which has been started
in the determination run.

28.5

Conditioning a titration cell

Manual control

▶ Conditioning

In manual control, you have the option of permanently conditioning one
titration cell for Karl Fischer titrations while carrying out a Karl Fischer
titration in a second titration cell.

Proceed as follows:

1 Create a method

■

Create a Karl Fischer method.

■

Save the method.

2 Load the method

■

Select the memory location where the desired method is saved.

■

Tap on the input field File name.
The method table with the saved methods is opened.

■

Select the desired method.

3 Start the conditioning

■

Tap on [Start].

The working medium is titrated to the endpoint. The status is kept
stable.