5 custom calibration buffers – Metrohm 900 Touch Control Manual User Manual
Page 174

15.5 Custom calibration buffers
■■■■■■■■■■■■■■■■■■■■■■
158
■■■■■■■■
900 Touch Control
15.5
Custom calibration buffers
Main dialog: System
▶ Templates ▶ Custom calib. buffers
In the Templates / Custom calibration buffers dialog, you can define
a buffer series of your own for pH calibrations with a maximum of five cal-
ibration buffers. Automatic buffer detection is active for this buffer.
Buffer 1 pH
pH value of the first calibration buffer.
The values that are actually measurable may vary from the values specified
here, depending on the hardware. The binding measurement value ranges
can be found in the "Technical specifications" chapter for the control
device used.
Input range
–20.000 - 20.000
Selection
off
Default value
off
Buffer 2...5 pH
See Buffer 1 pH.
[Temp. table]
Display the temperature table for the selected buffer, see following chap-
ter.