2 stop conditions – Metrohm 900 Touch Control Manual User Manual
Page 467

■■■■■■■■■■■■■■■■■■■■■■
29 Parameters
900 Touch Control
■■■■■■■■
451
29.15.5.2
Stop conditions
The conditions for canceling the dosing are defined under [Stop condi-
tions].
Stop volume
The dosing is stopped when the specified volume has been dosed since
the start of the dosing. This volume should be adjusted to the size of the
titration vessel in order to prevent the contents from running over.
Input range
0.00000 - 99999.9 mL
Selection
off
Default value
off
Stop time
The dosing is stopped when the specified time has elapsed since the start
of the dosing.
Input range
0 - 999999 s
Selection
off
Default value
off
Filling rate
Rate at which the dosing cylinder is filled.
Input range
0.01 - 166.00 mL/min
Selection
maximum
Default value
maximum
maximum
The maximum filling rate depends on the cylinder volume (see Chapter
32, page 488).