Chapter 6. Data Formatting and Reports

93

Each custom report template is released as a controlled document with its own part

number and revision level. All validation procedures and results will be included with
each release.

The second section contains the validation results for the positive and negative
controls, as specified in the Analysis Settings worksheet. Included in the control
validation results is the number of valid controls and mean absorbance value for each
control group (positive and negative). These values are highlighted in red if the
control group has failed validation.

Figure 70. Control validation.

The third section of the Report worksheet contains the calculated cutoff values for
identifying positive, negative, and indeterminate samples, as specified in the Analysis
Settings worksheet.

Figure 71. Cutoff values.

The forth section contains the test results, laid out in a 96-well plate format. Positive
control wells (PC1 and PC2) and negative control wells (NC1 through NC4) show
the reading absorbance values. Controls that don’t pass validation are flagged in red
with strikethrough marks.