Microplate Manager User Guide

40

Enter the number of readings you want to take in the Readings column.

Next enter the interval between the start of two consecutive readings in the Interval
column. Note that this is the minimum time necessary for one complete reading plus
any mixing time you specify. For example, when using the Ultramark with a standard
12

× 8 (96-well) plate, the smallest interval you can specify is 7 seconds. For custom

plate sizes, the smallest interval you can specify is 450 seconds (7.5 minutes). If the
interval you enter is too short, a message box will tell you the minimum interval time.

Finally, you can specify a mixing time for your plate. If no mixing is desired, enter a
0.

If you specify a mixing time, be sure to include this time in the interval between

readings (Interval column).

Analysis Options

To access the other Kinetic Protocol options, click on the Analysis Options... button.
This will open the Analysis Options dialog box.