Microplate Manager User Guide

82

Figure 61. Control Report.

The Control Report displays absorbance or velocity data for each well as a
percentage of the range between the positive and negative controls in an assay.

If no positive control is defined for the assay, the highest absorbance value or highest
velocity among the wells defined as unknowns or standards is used. If no negative
control is defined, then 0.000 is used as the negative control.

Standard Curve Report

The Standard Curve Report displays the absorbance (OD) values of the standards
plotted against their concentrations (see page 70 for information on entering standard
concentrations). Using the report, you can select the best regression type and scaling
method for generating a standard curve for your data.

With the Data File open, select Standard Curve Report from the VIEW menu.